Study the quality assurance of conventional
x-ray machines at medical city in Baghdad |
|
|
The primary goal of a radiology quality assurance (QA) program
is to ensure adequate clinical diagnostic information and low
dose (ALARA principle). This can be achieved by optimum
operating parameters such as kVp (peak voltage the highest kilo
voltage used in producing a radiograph), Entrance skin dose ESD,
linearity of tube current and exposure time and beam quality
assess of x-ray machine. Six conventional X-ray instruments in
medical city at Baghdad from different origins and different
manufacturing time. Protocol for QA was perform to measure
different parameter such as patient entrance surface dose air
kerma ESD, reproducibility of peak tube voltage, linearity of
tube current, linearity of exposure time, and the filtration.
The mean (ESD) were measured for different organ using (Dosimax),
(Unfors) ionization chamber and thermoluminance dosimeter (TLD).
(TLD) reading was higher than average value of ionization
chamber because the scattering of radiation inpatient body, the
scattering coefficient were range between (1.1 – 1.05). The
voltage stability for the machine were tested for three machines
(A, B, and C) and found the percentage of voltage ripple to
machine (B and C) is (3.79) and (4.856) respectively and for
machines A was 11.4. The Linearity of current and exposure time
exposure were assess, and found that this relationship is linear
for the machines (B and C) but not for the machine (A). Half
Value Layer (HVL) test for the three machines was measured.
Keywords: QA, diagnostic x-ray, kVp, current linearity
|
|